Infection ambulance is specially designed for the transporta-tion of infected patients (who have contagious diseases ilke COVID-19, BSL-4, Ebola, SARS, Mers, Avian Flu, Swine Flu (H110) in the best pos-sible high isolation.
Infection ambulance is specially designed for the transporta-tion of infected patients (who have contagious diseases ilke BSL-4, Ebola, SARS, Mers, Avian Flu, Swine Flu (H110) in the best pos-sible high isolation.
Negative pressure and desired air circulation is created in the patient compartment by use of air-sucking-fans. Circulated air is passed through the HEPA filters to decontaminate the air used in the compartment and some of it blowed out of the ambulance white some of it blowed through the compartment again as clean air. So it is provided that the infected patient is carried in safe with no harm to outside environment and the ambulance crew by use of negative pressure and the desired air circulation. Desired air circulation is that the air blowed into the compartment dominate the rising air from the flooring and that sucked by the fans so that always decontaminated air is present just above the patient.
Patient compartment pressure can be lowered to - 260 pas-cal forcing the contaminated air move through the filters and so that it is provided best possible air circulation creating maximum safety inside the compartment.
Ideally, the pressure inside the compartment can be ad-justed to - 100 Pascal.
For ensuring the Tong lasting of hygenic environment, the number of 0.3 micron size particules is kept below 2.000.000. ( In tests done in the development phase, it is observed that the number of particules of 0.3 micron is around 10.200)
Two echelon filtration (F7-H13 HEPA filters) system is used for the air circulation inside the compartment.
Airflow rate of the HVAC system used in the compartment for air circulation is max.1800 m3/h and can be adjusted to the required values. With the use of high tech otomation systems, temperature, humidity, pressure and the particule parameters can be adjusted by the crew responsible.
The risk of spreading the contaminated air outside is eliminated by using specially designed EMS Guard Filters. Decontamination rate of 99.95 Z, is reached with the use of F9 and H13 filters.
There are 2 filter of Guard HEPA filters in the left ant anf right side of the compartment.
Cooling and heating systems are protected against bac-teria and viruses by using HEPA filter systems.
All the material and equipment used inside the compartment have rounded corners for easy de-contamination and are Tong lasting products.
Epoxy floor covering is special for preventing bacteria and virus population growing and easy to decontaminate. It has tong lasting usage life and not affected from decontamination solutions (Formaldehit-Paraformaldehit-H202)
The patient compartment is designed to comply with the requirements of TS EN 1789 standard.
All the standard ambulance medikal equipments are located in patient cabins.